8200-M585-O1 Rev D 210 Installation & Technical 38
SETUP AND CALIBRATION, CONT.
b= (Break Range)
Press the ENTER key to show the current setting for the break range. The break range is a
number from 1 to 255 that corresponds to the number of division change to break out of the
filtering.
Press the ENTER key to keep the displayed value or use the numeric keys to enter a new
value and press the ENTER key to save the new setting. Note that entry of a 0 disables this
feature.
h= (High Frequency Filter Value)
Press the ENTER key to show the current setting for the high frequency filter level. The filter
level is a number from 0 to 99 that corresponds to the level of filtering with 99 being the
greatest filtering and 1 the least.
To accept the value displayed, press the ENTER key, otherwise, use the numeric keys to enter
a new value then press the ENTER key to save it. Note that entry of a 0 disables this feature.
P= (Preliminary Filter)
Press the ENTER key to show the current setting for the preliminary filter. The preliminary
filter is a number from 0 to 99 that corresponds to the number of weight samples that will be
filtered without breaking (regardless of what b= is set to). This filter is ideal for fast paced
Filling operations.
To accept the value displayed, press the ENTER key, otherwise, use the numeric keys to enter
a new value and then press the ENTER key to save it. Note that entry of a 0 or 1 disables this
feature.
Sr= (Sample Rate)
Press the ENTER key to show the current setting for the sample rate. The value displayed is
the sample rate in samples per second.
Press the ENTER key to save the displayed value or use the numeric keys to enter a new
value (1 to 100) and press the ENTER key to save it.
UnS= (Motion Range)
Press the ENTER key to view the current setting for the range of motion detection. If the
displayed value is acceptable, press the ENTER key to save it. Otherwise, use the numeric
keys to enter the new range (the number of divisions of change permitted before indicating
unstable), then press the ENTER key to save the new setting. Allowable range values are: 0
through 99 divisions.
SC= (Stable Count)
Press the ENTER key to view the current setting for the number of consecutive stable weight
readings before indicating stable weight. This helps filter weight readings for stability for use
with Auto Print on Balance, or and anything trying to capture stable weight.
If the displayed value is acceptable, press the ENTER key to save it. Otherwise, use the
numeric keys to enter a new value and press the ENTER key to save the new setting.
Allowable values for the stable count are: 3 through 255.