8200-M585-O1 Rev D 210 Installation & Technical 77
TROUBLESHOOTING, CONT.
Diagnostic Test Mode
To Start Diagnostic Test Mode
1. Remove calibration access screw.
2. Press ON/OFF key to turn indicator ON.
3. Insert a small screwdriver or other tool into the access hole and press the CAL switch.
4. Display will change to SETUP.
5. Press UNITS/TEST key.
6. Display will change to tESt=.
7. Press ENTER key to show current test function value (which will be 0, test feature off).
8. Using the numeric keys, enter test function value and then press ENTER key.
4 = raw A/D readings
16 = display milliVolts
9. Display will change to SETUP.
10. Press key to begin test.
11. Display will change to show test data.
To Change Diagnostic Test Mode
1. With indicator displaying test data, insert a small screwdriver or other tool into the
access hole and press the CAL switch.
2. Display will change to SETUP.
3. Press UNITS/TEST key.
4. Display will change to tESt=.
5. Press ENTER key to show current test function value.
6. Using the numeric keys, enter test function value and then press ENTER key.
4 = raw A/D readings
16 = display milliVolts
7. Display will change to SETUP.
8. Press key to begin test.
9. Display will change to show test data.
To Stop Diagnostic Test Mode
1. With indicator displaying test data, insert a small screwdriver or other tool into the
access hole and press the CAL switch.
2. Display will change to SETUP.
3. Press UNITS/TEST key.
4. Display will change to tESt=.
5. Press ENTER key to show current test function value.
6. Using the numeric keys, enter a 0 for the test function value (0 = test feature off) and
then press ENTER key.
7. Display will change to SETUP.
8. Press key to stop test and reset indicator.
9. The indicator is now ready for normal operation.
IMPORTANT! Once Diagnostic Test Mode has been enabled, the indicator will
power on with TEST on the display. Operator must press ENTER key to stay in
Diagnostic Test Mode or press key to disable Diagnostic Test Mode.