8200-M698-O1 Rev B 225 Installation and Technical
72
CALIBRATION AND SETUP, CONT.
1. FILTER=X (DIGITAL FILTERING)
With SCALE 1 SETUP MENU #2 displayed the current setting for the FILTER=
parameter will be shown. Note that X is the current value for the digital filtering. If the
setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enter the
value for the digital filtering and then press the ENTER key to save the new setting and
return to SCALE 1 SETUP MENU #2. Allowable values are 0, 1, 2, or 3.
0 DISABLED (NO Filtering)
1 MINIMAL FILTERING (Sample Rate = 2)
2 MODERATE FILTERING (Sample Rate = 1)
3 CUSTOM FILTERING (additional parameters available)
2. FMAX=XXX (MAXIMUM FILTER LEVEL)
With SCALE 1 SETUP MENU #2 displayed the current setting for the FMAX=
parameter will be shown. Note that XXX is the current value for the maximum filter
level. The filter level is a number from 1 to 255 that corresponds to the level of filtering
with 255 being the greatest filtering and 1 the least. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 2 key, the ENTER key and then using the numeric keys, enter the
value for the new maximum filter level and then press the ENTER key to save the new
setting and return to SCALE 1 SETUP MENU #2. Allowable values are 1 through 255.
3. FBRK=XXX (FILTER BREAK RANGE)
With SCALE 1 SETUP MENU #2 displayed the current setting for the FBRK=
parameter will be shown. Note that XXX is the current value for the filter break range.
The break range is a number from 1 to 255 that corresponds to the number of division
changes to break out of the filtering. If the setting displayed is acceptable, proceed to
the next setup parameter.
Otherwise, press the 3 key, the ENTER key and then using the numeric keys, enter the
value for the new filter break range and then press the ENTER key to save the new
setting and return to SCALE 1 SETUP MENU #2. Allowable values are 1 through 255.