8200-M698-O1 Rev B 225 Installation and Technical
71
CALIBRATION AND SETUP, CONT.
9. WT INTERVAL=X (WEIGHT INTERVAL “SINGLE OR DUAL”), CONT.
CAUTION! When using the dual interval feature of the 225, do not attempt to use
a division value in the low range that is too small to provide adequate signal
strength for stable operation of the indicator. Most scale installations will allow the
smaller division to be 1/2 of the larger division value when the dual interval feature
is used.
The inherent sensitivity and capabilities of the 225 will allow you to select
combinations that are beyond the practical limits of today’s load cells for
dependable, stable performance. The appearance of such a situation will be
instability in the lower range, while the higher range will operate, as it should.
10. TYPE=XXXXXXXX (SELECT SCALE TYPE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the
TYPE= parameter will be shown. Note that XXXXXXXX is the current value (scale
type). If the setting displayed is acceptable, press EXIT (Navigation Keys UP Arrow)
to return to the SETUP MENU #2.
Otherwise, press the 1 and 0 keys, the ENTER key and then using the numeric keys,
enter the value for the scale type and then press the ENTER key to save the new
setting and return to the SCALE 1 SETUP MENU #1. Allowable values are 0, 1 or 2.
0 = ANALOG 1 = GUARDIAN 2 = SERIAL
NOTE: To ensure proper operation, the settings in SCALE SETUP MENU #2 MUST
match the settings of the local indicator.
With SCALE 1 SETUP MENU #1 displayed, press NEXT (Navigation Keys Down Arrow)
to proceed to SCALE 1 SETUP MENU #2.
SCALE 1 SETUP MENU #2
1. FILTER=X 6. CAP=XXXXXXX
2. FMAX=XXX
3. FBRK=XXX
4. INTERVAL=XX
5. DPP=X
Enter Selection: 0 ` NEXT ^ EXIT
NOTE: Although shown on the scale setup menu, the setup parameters FMAX=
and FBRK= can only be selected when CUSTOM FILTERING (FILTER=3)
has been selected.