6. SAMPLES=XXX (SAMPLE RATE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the SAMPLES=
parameter will be shown. Note that XXX is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 6 key, the ENTER key and then using the numeric keys, enter the
value for the sample rate in samples per second and then press the ENTER key to save
the new setting and return to SCALE 1 SETUP MENU #1. Allowable values for the
sample rate are 1 through 100.
7. STABLE=XX (STABLE “MOTION RANGE”)
With SCALE 1 SETUP MENU #1 displayed the current setting for the STABLE=
parameter will be shown. Note that XXX is the current value for the range of motion
detection. If the setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 7 key, the ENTER key and then using the numeric keys, enter the
value for the new range (the number of divisions of change permitted before indicating
unstable) and then press the ENTER key to save the new setting and return to
SCALE 1 SETUP MENU #1. Allowable values are 0 through 99 divisions.
8. STBL CNT=XXX (STABLE COUNT “CONSECUTIVE STABLE READINGS”)
With SCALE 1 SETUP MENU #1 displayed the current setting for the STBL CNT=
parameter will be shown. Note that XXX is the current number of consecutive stable
weight readings before indicating stable weight. If the setting displayed is acceptable,
proceed to the next setup parameter.
Otherwise, press the 8 key, the ENTER key and then using the numeric keys, enter the
value for the stable count and then press the ENTER key to save the new setting and
return to SCALE 1 SETUP MENU #1. Allowable values are 1 through 255.
9. WT INTERVAL=X (WEIGHT INTERVAL “SINGLE OR DUAL”)
With SCALE 1 SETUP MENU #1 displayed the current setting for the
WT INTERVAL= parameter will be shown. Note that X is the current value. If the
setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 9 key, the ENTER key and then using the numeric keys, enter the
value for the weight interval and then press the ENTER key to save the new setting and
return to SCALE 1 SETUP MENU #1. Allowable values are 1 or 2.
WT INTERVAL=1 WT INTERVAL=2
Single Interval (One weight range) Dual Interval (Two weight ranges)