Fusion 200
1. If necessary, move and tighten the infusion safety collar (see Adjusting the Safety
Collar).
2. While pressing in the block release button, slide the pusher block all the way to the left.
3. Loosen the screw knobs for the flange retaining bracket and adjust the bracket such that
the flange of the syringe barrel will fit in between. Do not tighten the bracket, yet.
4. Pull up on the spring-loaded syringe clamp and place the syringe in one of the channels
of the holder block. Ensure that the flange of the syringe barrel is flush against the edge
of the holder block. Slowly lower the syringe clamp to secure the syringe in place.
5. Loosen the bracket clamp knob for the plunger retaining bracket and adjust the gap such
that the plunger cap of the syringe may fit.
6. Press the block release button and slide the pusher block to the right until it is flush
against the syringe plunger. The plunger cap for the syringe should fit in the gap
between the bracket and the pusher block.
7. Tighten the bracket clamp knob for the plunger retaining bracket to secure the syringe
plunger in place.
8. Move the flange retaining bracket flush against the flange of the syringe barrel and
tighten the bracket clamp knobs to secure the syringe barrel flange in place.