4 Operation
BodyGuard 595Color Vision Operator Manual
38
20.
In case of patient bolus: Patient triggers the
bolus through an external bolus cable.
Set bolus dose volume and press .
21.
Set a maximum bolus dose allowed for the
protocol. Press to continue.
22.
Set the lockout time (to prevent over dose by
patient) and press .
23.
Set the maximum lockout time allowed for the
protocol. Press to continue.
24.
In case of auto bolus: The patient will
periodically receive an automatic bolus until the
hour limit or VTBI have been reached.
Press key.
PCA 10:00 - 70%
A: MORPHINE
Warning ! Patient will
Periodically receive
an automatic bolus.
Press to Confirm
PCA 10:00 - 70%
A: MORPHINE
Patient Bolus Dose
3 mL
Enter and Press Key
PCA 10:00 - 70%
A: MORPHINE
Maximum Bolus Dose
3 mL
Enter and Press Key
PCA 10:00 - 70%
A: MORPHINE
Lockout Time
5 Minutes
Enter and Press Key
PCA 10:00 - 70%
A: MORPHINE
Maximum Lockout Time
1 Minutes
Enter and Press Key