4 Operation
BodyGuard 595Color Vision Operator Manual
37
17.
Set the protocol maximum rate. This will set the
maximum rate allowed for titration.
18.
If your protocol requires bolus press to
confirm.
19.
Set bolus type:
For patient bolus press .
For auto bolus press and then press
.
For patient bolus combined with auto
bolus press twice and then press .
NOTE: auto bolus option can be enabled or disabled from the Technician menu.
See Auto Bolus section.
PCA 10:00 - 70%
A: MORPHINE
Protocol Maximum Rate
100 mL/hr
Enter and Press Key
PCA 10:00 - 70%
A: MORPHINE
Bolus Required ?
Answer or
PCA 10:00 - 70%
A: MORPHINE
Patient Bolus ?
Answer or
PCA 10:00 - 70%
A: MORPHINE
Auto Bolus ?
Answer or
PCA 10:00 - 70%
A: MORPHINE
Patient Bolus
Combined with
Auto-Bolus ?
Answer or