BodyGuard 595™Color Vision Operator Manual
If the previous infusion was not over, the following
screen will appear:
To continue previous infusion, press ,
The screen will then prompt for code. Enter level
one access code and press . Review the
selected protocol details and press until the
pump prompts to "Start infusion?" or "Start Auto-
Bolus?". Confirm with and start infusion.
To start a new infusion press . The screen than will prompt for code. Enter
level one access code and press .
To prime the set select "Prime" and confirm with
. Prime as explained in the priming section
of this manual.
Warning: do not attach the set to the patient’s access device before priming the
administration set.
EPIDURAL 10:00 - 70% █
■
K: BUPIVICAINE
Press key to Resume
Press key for Menu
EPIDURAL 10:00 - 70% █
■
Start Infusion ?
Press key to start
EPIDURAL 10:00 - 70% █
■
Prime
Select Protocol
Change Bag
View Protocol Settings
View Shift Totals
View Patient History
View Event Log
Clinician Bolus
Select , and press
EPIDURAL 10:00 - 70% █
■
Start Auto-Bolus?
Press key to start