111
-
-
-
-
IS2 TOO LOW
SYS measuring value of channel 2 is
below lower alarm limit.
User-selectable
ID2 TOO HIGH
DIA measuring value of channel 2 is
above upper alarm limit.
User-selectable
ID2 TOO LOW
DIA measuring value of channel 2 is
below lower alarm limit.
User-selectable
IM2 TOO HIGH
M AP measuring value of channel 2 is
above upper alarm limit.
User-selectable
IM2 TOO LOW
M AP measuring value of channel 2 is
below lower alarm limit.
User-selectable
Technical alarms:
Message Cause
Alarm
Level
Remedy
IBP 1 SENSOR OFF
IBP cable of channel
1 fal i s off from
monitor.
LOW
Make sure that cable is properly
connected.
IBP 2 SENSOR OFF
IBP cable of channel
2 fa i ls off from
monitor.
LOW
Make sure that cable is properly
connected.
IBP (1,2) INIT ERR
IBP module failure HIGH
Stop using measuring function of
IBP module, notify biomedical
engineer or Our service staff.
IBP (1,2) INIT ERR 1
IBP (1,2) INIT ERR 2
IBP (1,2) INIT ERR 3
IBP (1,2) INIT ERR 4
IBP (1,2) INIT ERR5
IBP (1,2) INIT ERR 6
IBP (1,2) INIT ERR7
IBP (1,2) INIT ERR8
IBP (1,2) COMM
STOP
IBP (1,2) module
failure or
communication
failure
HIGH
Stop using measuring function of
IBP module, notify biomedical
engineer or Our service staff.
IBP(1,2) COMM ERR
IBP(1,2)
communication error
HIGH
Stop using measuring function of
IBP module, notify biomedical
engineer or Our service staff.
IBP1 ALM LMT ERR
Functional safety
failure
HIGH
Stop using measuring function of
IBP module, notify biomedical
engineer or Our service staff.
IBP2 ALM LMT ERR
Functional safety
failure
HIGH
Stop using measuring function of
IBP module, notify biomedical
engineer or Our service staff.
Prompt message (general alerts) :
Message Cause
Alarm
Level
IBP1 SYS EXCEED
Systolic measuring value of channel 1 is
beyond measurement range.
HIGH
IBP1 DIA EXCEED
Diastolic measuring value of channel 1 is
beyond measurement range.
HIGH
IBP1 MEAN EXCEED
Mean measuring value of channel 1 is
beyond measurement range.
HIGH