- 139 -
-
-
-
Appendix
Appendix
Appendix
Appendix II
II
II
II
System
System
System
System Alarm
Alarm
Alarm
Alarm Prompt
Prompt
Prompt
Prompt
PROMPT
PROMPT
PROMPT
PROMPT CAUSE
CAUSE
CAUSE
CAUSE MEASURE
MEASURE
MEASURE
MEASURE
"XX TOO HIGH "
XX value exceeds the higher
alarm limit.
Check if the alarm limits are
appropriate and the current situation of
the patient.
"XX TOO LOW "
XX value is below the lower
alarm limit.
XX represents the value of parameter such as HR , ST1, ST2, RR, SpO
2
,
IBP,
NIBP, etc in the system.
"ECG WEAK SIGNAL "
The ECG signal of the patient
is too small so that the system
can not perform ECG
analysis.
Check if the electrodes and lead
wires are connected correctly and the
current situation of the patient.
NO PULSE
The pulse signal of the patient
is too small so that the system
can not perform pulse
analysis.
Check the connection of the sensor
and the current situation of the
patient.
" RESP APNEA "
The respiration signal of the
patient is too small so that the
system cannot perform RESP
analysis.
Check the connection of the linking
wire and the current situation of the
patient.
" CO
2
APNEA "
The respiration signal of the
patient is too small so that the
system cannot perform RESP
analysis.
Check the connection of CO
2
sensor
and the current situation of the
patient.
" ASYSTOLE "
Patient suffers from Arr. Of
ASYSTOLE.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
" VFIB/VTAC "
Patient suffers from Arr. of
VFIB/VTAC.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
" COUPLET "
Patient suffers from Arr. of
COUPLET.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
" BIGEMINY "
Patient suffers from Arr. Of
BIGEMINY.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
" TR I GEMINY "
Patient suffers from Arr. of
TR I GEMINY.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
" R ON T "
Patient suffers from Arr. of R
ON
T.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
"PVC"
Patient suffers from Arr. of
PVC.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
"TACHY" Patient suffers from TACHY.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
" BRADY" Patient suffers from BRADY.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours