- 140 -
-
-
-
"VT>2"
Patient suffers from Arr. of
VT >2 .
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
MISSED BEATS
Patient suffers from Arr. of
MISSED BEATS.
Check the current situation of the
patient. Check the connection of the
electrodes and lead wires.
"PNP" The pacemaker is not paced.
Check the connection of the
pacemaker.
Check the connection of electrodes
and lead wires. Check the current
situation of the patient.
"PNC"
No pacemaker signal is
captured.
Check the connection of the
pacemaker.
Check the connection of electrodes
and lead wires. Check the current
situation of the patient.
"ECG LEAD OFF "
ECG lead is not connected
correctly.
Check the connection of ECG lead
wire.
"ECG V LEAD OFF ";
The V lead wire of ECG is not
connected correctly.
Check the connection of V lead wire.
"ECG LL LEAD OFF ";
The LL lead wire of ECG is
not connected correctly.
Check the connection of LL lead wire.
"ECG LA LEAD OFF ";
The LA lead wire of ECG is
not connected correctly.
Check the connection of LA lead wire.
"ECG RA LEAD OFF ";
The RA lead wire of ECG is
not connected correctly.
Check the connection of RA lead
wire.
"ECG C LEAD OFF ";
The C lead wire of ECG is not
connected correctly.
Check the connection of C lead wire.
"ECG F LEAD OFF";
The F lead wire of ECG is not
connected correctly.
Check the connection of F lead wire.
"ECG L LEAD OFF ";
The L lead wire of ECG is not
connected correctly.
Check the connection of L lead wire.
"ECG R LEAD OFF ";
The R lead wire of ECG is not
connected correctly.
Check the connection of R lead wire.
SPO
2
SENSOR OFF
SpO
2
sensor may be
disconnected from the
patient or the monitor.
Make sure that the monitor and the
patient are in correct connection with
the cables.
SPO
2
INIT ERR
SpO
2
module failure
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
SPO
2
INIT ERR 1
SPO
2
INIT ERR 2
SPO
2
INIT ERR 3
SPO
2
INIT ERR 4
SPO
2
INIT ERR 5
SPO
2
INIT ERR 6
SPO
2
INIT ERR 7
SPO
2
INIT ERR 8
SPO
2
COMM STOP
SpO
2
module failure or
communication error
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours