- 141 -
-
-
-
SPO
2
COMM ERR
SpO
2
module failure or
communication error
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
SPO
2
ALM L M T ERR Functional safety failure
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
PR ALM LMT ERR Functional safety failure
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
Alarm information:
SpO
2
NO SENSOR
Sensor not fully inserted into
the connector.
May be an incorrect sensor, or a
defective sensor or cable. Insert
sensor into the connector. Disconnect
and reconnect sensor. Refer to the
instructions for the sensor being
used.
Sensor inserted upside down.
Disconnect and reconnect he sensor
with the logos matching.
SpO
2
SENSOR
OFF
SpO
2
sensor may be
disconnected from the patient
or the monitor.
Disconnect and reconnect the sensor.
Reattach sensor.
SpO
2
SENSOR FAULT
This message appears when
the sensor is faulty
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
SpO
2
UNRECOGNIZED
SENSOR
board does not recognize the
sensor.
Make sure that the monitor and the
patient are in correct connection with
the cables.
SpO
2
INCOMPATIBLE
SENSOR
This message is displayed
when the sensor is finding
incompatible sensor.
Make sure that the monitor use
incompatible sensor.
SpO
2
INTERFERENCE
Outside signal or energy
preventing reading.
Remove outside interference.
SpO
2
PULSE SEARCH
Unit is searching for the
patients pulse.
If values are not displayed within 30
seconds, disconnect and reconnect
sensor. If pulse search continues,
remove sensor and replace on a
better perfused site.
SpO
2
LOW
PERFUSTION
Signal too small. Move sensor to better perfused site.
SpO
2
TOO
MUCH LIGHT
Too
much light on
patient(sensor). Inadequate
tissue covering sensor
detector.
Remove or reduce lighting. Cover
sensor from light.
Reposition sensor.
SpO
2
LOW
SIGNAL IQ
Low signal quality.
Ensure proper sensor application.
Mover sensor to a better perfused
site.
SpO
2
This message appears when
the Set board malfunctions.
Stop using the measuring function of
SpO
2
module, notify biomedical