When monitoring a patient, you may freeze the waveforms of interest so as to view them carefully.
Generally you can review maximally 3 4 seconds of a frozen waveform. If required, you may also use
recorder to print out a frozen waveform. The Freeze function of this monitor has following features:
Freeze status can be activated on any operating screen;
At the same time of entering the Freeze status, the system exits all other operating menus.
Besides, the system freezes all waveforms in the Waveform area of the Basic Screen, or
Full-lead ECG waveforms and the extra waveform (if available) on the Full-lead ECG screen.
Nevertheless the Parameter area refreshes normally.
In the Freeze status, it does not affect the display and refresh of the Trend Graph area on the
trend screen, the display and refresh of oxyCRG on the Dynamic Refresh screen, or the
display and refresh of the View B ed window on the View B ed screen .
The frozen waveforms can be reviewed or recorded.
5.2
5.2
5.2
5.2 Enter/Exit
Enter/Exit
Enter/Exit
Enter/Exit Freeze
Freeze
Freeze
Freeze Status
Status
Status
Status
Enter
Enter
Enter
Enter Freeze
Freeze
Freeze
Freeze Status
Status
Status
Status
In the Non-Freeze status, press the FREEZE button on the front panel of the monitor to let the system
exit the Menu being currently displayed (if available), then enter the Freeze status and display the
popup FREEZE menu. In the Freeze status, except View B ed waveforms, all other waveforms are
frozen. In other words, the system will no longer refresh all other waveforms.
Exit
Exit
Exit
Exit Freeze
Freeze
Freeze
Freeze Status
Status
Status
Status
In the Freeze status, executing any of the following operations will command the system to exit the
Freeze status:
Select the EXIT option on the FREEZE menu;
Press the FREEZE button on the front panel again;
Press the non-immediate-to-execute button (such as a button once pressed, a menu will pop
up for you to further select an option )on the front panel and system buttons of MAIN and
MENU;
Execute any operation that may trigger the adjustment of the screen or display of a new menu.
After exiting the Freeze status, the system will discharge the Freeze status, clear screen
waveforms and resume to display real-time waveforms.

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours