- 142 -
-
-
-
BOARD FAULT
engineer or our service staff.
SpO
2
COMMUNICATION
ERROR
This message is displayed
when the front end module is
having problems
communicating ( ie: framing
errors or bad checksums)
with the board.
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
SpO
2
COMMUNICATION
STOP
This message is displayed
when the host can not receive
the data from board for 5
seconds
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
SpO
2
INIT ERR
This message is displayed
when the SpO
2
module
initialization error happened.
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
"TEMP1 SENSOR OFF "
TEMP1 sensor is not
connected correctly.
Check the connection of TEMP1
sensor.
"TEMP2 SENSOR OFF "
TEMP2 sensor is not
connected correctly.
Check the connection of TEMP2
sensor.
"TEMP1 SENSOR OFF "
TEMP1 sensor is not
connected correctly.
Check the connection of TEMP1
sensor.
"TEMP2 SENSOR OFF "
TEMP2 sensor is not
connected correctly.
Check the connection of TEMP2
sensor.
"IBP1 LEAD OFF "
IBP1 sensor is not connected
correctly.
Check the connection of IBP1 sensor.
"IBP2 LEAD OFF "
IBP2 sensor is not connected
correctly.
Check the connection of IBP2 sensor.
" IBP1 NEED ZERO-CAL "
Zero calibrating must be done
before measuring in IBP1
Do zero calibrating for IBP1
" IBP 2 NEED ZERO-CAL "
Zero calibrating must be done
before measuring in IBP2
Do zero calibrating for IBP2
" TB SENSOR OFF "
TB sensor is not connected
correctly.
Check the connection of TB sensor.
"ECG NOISE "
Rather large interference
signals appear in the ECG
signals.
Check the connection of ECG lead
wire. Check the current situation of
the patient. Check if the patient
moves a lot.
"XX INIT ERR X"
XX has error X during
initialization.
Re-start up the monitor or re-plug
in/out the module. If the error still
exists, contact the manufacturer.
"XX COMM STOP "
XX cannot communicate with
the host.
"XX COMM ERR "
XX cannot communicate
normally with the host.
XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, IBPmodule, etc.
"XX ALM LMT ERR "
The alarm limit of XX
parameter is modified by
chance.
Contact the manufacturer for repair.
"XX RANGE EXCEEDED "
The measured value of XX
parameter has exceeded the
measuring range of the
system.
Contact the manufacturer for repair.
XX represents the parameter name in the system such as HR, ST1, ST2, RR, SpO
2
,
IBP,
NIBP, etc.

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours