- 143 -
-
-
-
" CO
2
Sensor Faulty" The Sensor Source Current Failure
Check that the sensor is properly
plugged in. Reinsert or reseat the
sensor if necessary. If error
persists,return sensor to factory
for servicing.
"CO
2
Sensor Over temp"
The sensor temperature is greater
than 40
Make sure sensor is not exposed
to extreme heat. If error persists,
return sensor to factory for
servicing.
"CO
2
Check Sampling Line"
This error occurs whenever the
pneumatic pressure is outside the
expected rang
Check that the sampling line is
not occluded or kinked.
"CO
2
Zero Error" An error was found during Zero
To
clear, check airway adapter
and clean if necessary. If this
does not correct the error,
perform an adapter zero.
"CO
2
Out of Range"
The value being calculated is
greater than the upper CO
2
limit.
If error persists, perform a zero.
"CO
2
Check Airway Adapter"
Usually caused when the airway
adapter is removed from the
sensor or when there is an optical
blockage on the windows of the
airway adapter. May also be
caused by failure to perform Zero
to When adapter type is changed.
To
clear, clean airway adapter if
mucus or moisture is seen. If the
adapter is clean, perform a Zero.
"CO
2
not initialized"
Barometric Pressure or gas
compensatins have not been set
since power on.
Set the Barometric Pressure and
gas compensations to clear this
error.
" REAL CLOCK NEEDSET "
When the system displays
2000-1-1, the system gives this
prompt reminding the user that the
current system time is not right.
Re-set up the system time. It is
better to set up the time just after
the start-up and prior to
monitoring the patient. After
modifying the time, the user had
better re-start up the monitor to
avoid storing error time.
" REAL CLOCK NOT EXIST "
The system has no cell battery or
the battery has run out of the
capacity.
Install or replace the
rechargeable battery.
" SYSTEM WD FAILURE "
The system has serious error.
Re-start up the system. If the
failure still exists, contact the
manufacturer.
" SYSTEM SOFTWARE
ERR "
" SYSTEM CMOS FULL "
" SYSTEM CMOS ERR "
" SYSTEM EPGA FAILURE "
" SYSTEM FAILURE 2"
" SYSTEM FAILURE 3"
" SYSTEM FAILURE 4"
" SYSTEM FAILURE 5"
"SYSTEM FAILURE6"
" SYSTEM FAILURE 7"
" SYSTEM FAILURE 8"
" SYSTEM FAILURE 9"
" SYSTEM FAILURE 10"
" SYSTEM FAILURE 11"
" SYSTEM FAILURE 12"
" KEYBOARD NOT The keys on the keyboard cannot Check the keys to see whether it

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours