- 144 -
-
-
-
AVAILABLE "; be used. is pressed manually or by other
object. If the key is not pressed
abnormally, contact the
manufacturer for repair.
" KEYBOARD COMM ERR ";
The keyboard has failure, which
cannot be used.
Contact the manufacturer for
repair.
" KEBOARD ERROR ";
" KEYBOARD ERR 1";
" KEYBOARD ERR 2";
" NET INIT ERR (G.)"
The network part in the system has
failure. The system cannot be
linked to the net.
Contact the manufacturer for
repair.
" NET INIT ERR (Ram)"
" NET INIT ERR (Reg)"
" NET INIT ERR (Mii)"
" NET INIT ERR (Loop)"
" NET ERR (Run1)"
"NET ERR(Run2)"
" NET ERR (Run3)"
"5V TOO HIGH "
The power part of the system has
failure.
If the prompt appears repeatedly,
contact the manufacturer for
repair.
"5V TOO LOW "
" POWER ERR 3"
" POWER ERR 4"
"12V TOO HIGH "
"12V TOO LOW "
" POWER ERR 7"
" POWER ERR 8"
"3.3V TOO HIGH "
"3.3V TOO LOW "
" CELL BAT TOO H I GH " Cell battery has problem.
Replace the battery. If the failure
still exists, contact the
manufacturer.
" CELL BAT TOO LOW "
The cell battery has low capacity or
the cell battery is not installed or
the connection is loose.
" RECORDER SELFTEST
ERR "
During the selftest, the system fails
connecting with the recorder
module.
Execute Clear Record Task
function in the recorder setup
menu to re-connect the host and
the recorder. If the failure still
exists, contact the manufacturer
for repair.
" RECORDER VLT HIGH "
The recorder module has voltage
failure.
Contact the manufacturer for
repair.
" RECORDER VLT LOW "
" RECORDER HEAD HOT "
The continuous recording time may
be too long.
After the recorder becomes cool,
use the recorder for output again.
If the failure still exists, contact
the manufacturer for repair.
" REC HEAD IN WRONG
POSITION "
The handle for pressing the paper
is not pressed down.
Press down the recorder handle
for pressing the paper.
" RECORDER OUT OF
PAPER "
No paper is in the recorder. Place the paper into the recorder.
" RECORDER PAPER JAM "
The paper in the recorder is
jammed.
Place the recorder correctly and
try again.
" RECORDER COMM ERR "
The communication of the recorder
is abnormal.
I n the recorder setup menu,
execute the function of clearing
record task. The function can
make the host and the recorder
connect again. If the failure still
exists, contact the manufacturer
" RECORDERS. COMM
ERR "

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours