- 145 -
-
-
-
for repair.
" RECORDER PAPER
W.P.
"
The paper roll of the recorder is not
placed in the correction position.
Place the paper roll in the correct
position.
" REC NOT AVAILABLE "
Cannot communicate with the
recorder.
I n the recorder setup menu,
execute the function of clearing
record task. The function can
make the host and the recorder
connect again. If the failure still
exists, contact the manufacturer
for repair.
"NIBP INIT ERR "
NIBP initialization error
Execute the reset program in the
NIBP menu. If the failure still
exists, contact the manufacturer
for repair.
"NIBP SELFTEST ERR "
" NIBP ILLEGALLY RESET "
During NIBP measurement, illegal
reset occurs.
Check the airway of NIBP to see
if there are clogs. Then measure
again, if the failure still exists,
contact the manufacturer for
repair.
"NIBP COMM ERR "
The NIBP communication part has
problem.
Execute the reset program in the
NIBP menu. If the failure still
exists, contact the manufacturer
for repair.
" LOOSE CUFF "
The NIBP cuff is not connected
correctly.
Re-connect the NIBP cuff.
" AIR LEAK "
The NIBP cuff is not connected
correctly or there are leaks in the
airway.
Check the connection of each
part or replace with a new cuff. If
the failure still exists, contact the
manufacturer for repair.
" AIR PRESSURE ERROR "
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
Check the connection of each
part or replace with a new cuff. If
the failure still exists, contact the
manufacturer for repair.
" WEAK SIGNAL "
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
Check if the setup of patient type
is correct. Check the connection
of each part or replace with a
new cuff. If the failure still exists,
contact the manufacturer for
repair
" RANGE EXCEEDED "
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
Check the connection of each
part or replace with a new cuff. If
the failure still exists, contact the
manufacturer for repair.
" EXCESSIVE MOTION " The patient arm moves.
Check the connection of each
part and the patient situation.
Measure again, if the failure still
exists, contact the manufacturer
for repair.
" OVER PRESSURE " Perhaps folds exist in the airway.
Check for the smoothness in the
airway and patient situation.
Measure again, if the failure still
exists, contact the manufacturer
for repair.
" SIGNAL SATURATED "
Problem happens when measuring
the curve. The system cannot
Check the connection of each
part and the patient situation.