- 146 -
-
-
-
perform measurement, analysis or
calculation.
Measure again, if the failure still
exists, contact the manufacturer
for repair.
" NIBP TIME OUT "
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
Check the connection of each
part and the patient situation.
Measure again, if the failure still
exists, contact the manufacturer
for repair.
" CUFF TYPE ERR "
Perhaps the used cuff does not fit
the setup patient type.
Check if the patient type is set up
correctly. Check the connection
of each part or replace with a
new cuff. If the failure still exists,
contact the manufacturer for
repair.
" PNEUMATIC LEAK " NIBP airway has leaks.
Check the connection of each
part or replace with a new cuff. If
the failure still exists, contact the
manufacturer for repair.
" MEASURE FAIL "
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
Check the connection of each
part and the patient situation.
Measure again, if the failure still
exists, contact the manufacturer
for repair.
" NIBP SYSTEM FAILURE "
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
Check the connection of each
part and the patient situation.
Measure again, if the failure still
exists, contact the manufacturer
for repair.

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours