- 138 -
-
-
-
5 0ml/Min . ± 10 ml/Min .
Measuring range
CO
2
0~ 150 mmHg
INSCO
2
0~ 150 mmHg
AwRR 0 ~150 BPM
Resolution
CO
2
0. 1mmHg (0~69 mmHg )
0.25 mmHg (70~150 mmHg )
INSCO
2
0. 1mmHg (0~69 mmHg )
0.25 mmHg (70~150 mmHg )
Accuracy
CO
2
± 2mmHg 0~40mmHg
± 5% of reading 41~7 0 mmHg
± 8 % of reading 71 ~ 100 mmHg
± 10% of reading 101 ~ 150 mmHg
AwRR ± 1 rpm
Initialization Time
Mainstream Capnogram displayed in less than 15 seconds at an
ambient temperature of 25 ,full specifications within 2 minutes.
S idestream Capnogram displayed in less than 20 seconds at an
ambient temperature of 25 ,full specifications within 2 minutes.
Mainstream Rise Time
Less than 60ms-Adult Reusable or Single-Patient-Use Airway Adapter
Less than 60ms-Infant Reusable or Single-Patient-Use Airway Adapter
Actualization interval about 1 Sec
Sidestream Delay Time: 2~3Sec
Alarm range
CO
2
0~150 mmHg
InsCO
2
0~150 mmHg
AwRR 0 ~150 BPM
Suffocation Alarm Delay
AwRR 10~60 Sec.