- 137 -
-
-
-
Accuracy
Pressure
Maximum Mean error
±
5mmHg
Maximum Standard deviation
±
8mmHg
Overpressure Protection
Adult Mode 297
±
3 mmHg
Pediatric Mode 240
±
3 mmHg
Neonatal Mode 147
±
3 mmHg
2.1
2.1
2.1
2.1 0
0
0
0 SpO
SpO
SpO
SpO
2
2
2
2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy
70% ~ 100%
±
2 %
0% ~ 69% unspecified
Actualization interval about 1Sec.
Alarm Delay 10 Sec.
Pulse Rate
Measuring and Alarm Range
0~ 25 0 bpm
Resolution 1bpm
Accuracy
±
2 bp m
2.1
2.1
2.1
2.1 1
1
1
1 TEMPERATURE
TEMPERATURE
TEMPERATURE
TEMPERATURE
Channel 2
Measuring and Alarm Range 0 ~ 50 ℃
Resolution 0.1 ℃
Accuracy
±
0. 2 ℃
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec.
2.1
2.1
2.1
2.1 2
2
2
2 IBP
IBP
IBP
IBP
Channel 2
Label ART, PA, CVP, RAP,
LAP,
ICP,
P1, P2
Measuring and alarm range
-50 ~3 5 0mmHg
Press Sensor
Sensitivity 5 uV/V/mmHg
Impedance 300-3000 Ω
Resolution 1mmHg
Accuracy ± 2% or 1mmHg w hi ch great
Actualization Interval about 1 Sec
2.1
2.1
2.1
2.1 3
3
3
3 CO
CO
CO
CO
2
2
2
2
Method Infra-red Absorption Technique
Measuring mode MainStream and Sidestream
Side-stream mode sampling gas flow rate