- 136 -
-
-
-
Tachycardia ventricular : amplitude =2mV(p-v) , heart rate = 195bpm
Ampitude Average(s) Range (s)
× 1 4.2 4.1~4.3
× 0.5 11.9 11.5~12.1
× 2 3.0 2.9~3.1
2.
2.
2.
2. 8
8
8
8 RESP
RESP
RESP
RESP I
I
I
I RATION
RATION
RATION
RATION
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 M Ω
Measuring Impedance Range: 0.3~5.0 Ω
Base line Impedance Range: 0.1 K Ω – 2.5 K Ω
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy
±
2 rpm
Apnea Alarm 10 ~ 40 s
2.
2.
2.
2. 9
9
9
9 NIBP
NIBP
NIBP
NIBP
Method Oscillometric
Mode Manual, Auto, STAT
Measuring Interval in AUTO Mode
1, 2, 3, 4, 5, 10, 15, 30, 60, 90 ,120,180,240,480 Min
Measuring Period in
STAT
Mode 5 Min
Alarm Type SYS, DIA, MEAN
Measuring and alarm Range
Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 16 5 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg