- 76 -
-
-
-
VT>2
Withou t
pacemaker
3 < the number of cluster PVCs < 5 User-selectable
COUPLET
Without
pacemaker
2 consecutive PVCs User-selectable
BIGEMINY
Without
pacemaker
Vent Bigeminy
User-selectable
TRIGEMIN
Y
W ithout
pacemaker
Vent Trigeminy User-selectable
R ON T
Without
pacemaker
A
type of single PVC under the
condition that HR<100 , R-R interval is
less than 1/3 the average interval,
followed by a compensating pause of
1.25X the average R-R interval (the next
R wave advances onto the previous T
wave) .
User-selectable
PVC
Without
pacemaker
Single PVC s not belonging to the type
of above mentioned PVCs.
User-selectable
TACHY All patients
5 consecutive QRS complex , RR
interval is less than 500ms .
User-selectable
BRADY All patients
5 consecutive QRS complex , RR
interval is longer than 1.5s.
User-selectable
MISSED
BEATS
Without
pacemaker
When HR is less than 100 beats/min.,
no heart beat is tested during the period
1.75 times of the average RR interval;
or
W hen HR is larger than 100 beats/min.,
no beat is tested with 1 second.
User-selectable
PNP
With
pacemaker
No QRS complex and pacing pulse are
availabe during the period 1.75 times
of the average R - R interval (only
considering patients with pacemaker.)
User-selectable
PNC
With
pacemaker
When pacing pulse is available, no
QRS exists during the period 1.75 times
of the average RR interval (only
considering patients with pacemaker.)
User-selectable
Patient type:
A ll patients : refers to perform Arr.analysis on patients either with pacemakers or without pacemakers.
Without pacemaker : refers to perform Arr. Analysis only on the patients without pacemakers.
With pacemaker : refers to perform Arr. Analysis only on the patients with pacemakers.
Prompt message:
Message Cause Alarm Level
ARR LEARNING
The QRS template building
required for Arr. Analysis is in
process.
No alarm
N
N
N
N ote
ote
ote
ote
Arrhythmia
Arrhythmia
Arrhythmia
Arrhythmia name
name
name
name displays
displays
displays
displays in
in
in
in the
the
the
the Alarm
Alarm
Alarm
Alarm Message
Message
Message
Message Area.
Area.
Area.
Area.

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours