- 78 -
-
-
-
11.9.4
11.9.4
11.9.4
11.9.4 RESP
RESP
RESP
RESP menu
menu
menu
menu
RESP
RESP
RESP
RESP SETUP
SETUP
SETUP
SETUP Menu
Menu
Menu
Menu
Pick RESP hot key on the screen to call up the following menu:
Figure
Figure
Figure
Figure 11-16
11-16
11-16
11-16 RESP
RESP
RESP
RESP SETUP
SETUP
SETUP
SETUP Menu
Menu
Menu
Menu
RESP alarm setting
ALM: pick "ON" to enable prompt message and data record during the RESP alarm; pick "OFF" to
disable the alarm function, and there will be a beside RESP .
ALM REC: pick "ON" to enable report printing upon RESP alarm.
ALM LEV: selectable from HIGH, MED and LOW . Level HIGH represents the most serious case.
ALM HI: used to set up the upper alarm limit.
ALM LO: used to set up the lower alarm limit.
RESP alarm is activated when the respiration rate exceeds set ALM HI value or falls below ALM LO
value.
RESP alarm limits:
Max. RR HI M in . RR LO Step
RESP ADU 120 0 1
RESP NEO/PED 150 0 1
APNEA ALM: t o set the standard of judging an apnea case. It ranges from 10 to 40 seconds,
increases / decreases by 5 .
SWEEP : Available options for RESP SWEEP are 6.25, 12.5 and 25.0 mm/s.
WAVE AMP: The user may set up the displaying amplitude of the RESP waveform. The selections
are × 0.25, × 0.5, × 1, × 2.
DEFAULT: pick this item to access the RESP DEFAULT CONFIG dialog box, in which the user may
select whether the FACTORY DEFAULT CONFIG or the USER DEFAULT CONFIG is to be used.
After selecting any of the items and exiting the dialog box, the system will pop up the dialog box
asking for the user s confirmation
RESP
RESP
RESP
RESP Alarm
Alarm
Alarm
Alarm Message
Message
Message
Message
Among physiological alarms, those belonging to the type that the parameter has exceeded the limits
may activate the recorder to automatically output the parameters and related measured waveforms
when the alarms occur on the condition that the alarm record switch in the related menu is On.
Tables below describe the possible physiological alarms, technical alarms and prompt messages