Scanning Electron Microscope EM-30PLUS Operating Manual
45
Molding material
If it is hard to cut specimen, first mold, cut, polish and
analyze it. For molding specimens, ion coating is necessary.
Glass material
When you cut it, you can see its cross section and tilt image
can be acquired using tilt holder. For glass specimens, ion
coating is essential.
Powder Material
Sprinkle the powder material on carbon tape and blow it with
blower so that it may not get into the equipment.
When the powder has not been removed with blower, it can
be the cause of equipment deterioration.