5.7.1 Precautions for Using BSE Sensor
Surface contamination and damage of BSE sensor
Contamination and damage of sensor and cables can have bad effect on the detection
of Signal. Be sure to have paid A/S service for damage caused by user negligence. Be
sure to use BSE Sensor with caution.
Precautions for rotation of BSE
BSE sensor requires caution for operation as there is a risk of damage when excessive
power is applied to the sensor during rotation.
(Rotation must be executed by the person who completed training on rotation only.)
Precautions for use of Multi Holder
When using Multi Holder, if Stub is projected upwards severely above Housing
Cover(WD<7mm), there can be a collision of the specimen with BSE sensor. (D=7mm
when Stub and Housing Cover are on the level.)
WD 7mm
WD 10mm
Excessive use of
WD(Example)
BSE sensor is structurally vulnerable to contamination and damage. It is prohibited to be
used by the user who had no official training or full knowledge on the manual.