What to do if difference between constants is > 30% during conductivity calibration on Crison Multimeter?
- EElizabeth HillSep 13, 2025
If, during conductivity calibration, the difference between the constants is greater than 30% on your Crison Multimeter, insert the probe into the appropriate standard and read again. Examine the probe and replace it if necessary.