SERVICE MANUAL
Pre-installation
(Rev. 0) RTG PLUS (120V)
5-3
5.2
5.25.2
5.2
Environmental condition
Environmental conditionEnvironmental condition
Environmental condition
The performance of the equipment are assured if the room environment
must be treated in such a way as to ensure a relative humidity between
50% and 75 % and a temperature between +50°F ÷ +104°F (+10° ÷ +40°).
WARNING:
Proper backing is absolutely essential for all wall mounted versions. The
wall bracket supplied must be attached to the wall bearing in mind that
the wall backing has a capacity to sustain a load of 2000N (about
441 pounds - 200kg) per bolt applied. Backing and proper support is the
full responsibility of the owner.
5.3
5.35.3
5.3
Unpacking
UnpackingUnpacking
Unpacking
ROTOGRAPH PLUS is shipped packed in two cartons when the unit is
CEPH ready and three cartons when the Ceph (Teleradiography) device is
included.
Dimensions, net weight, gross weight and the packing contents are listed
in the following table:
Weight
Packing dimension Contents
Net Gross
40x37½x29½ inches
(100x95x75 cm)
- Motor group
- Tubehead
- Counterweights
- Accessories
331
pounds
(150 kg)
364
pounds
(165 kg)
98½x8¾x7 inches
(240x22x30 cm)
- Column
88
pounds
(40 kg)
99
pounds
(45 kg)
57x19¾x23¾ inches
(148x64x64 cm)
- Ceph arm
- Accessories
51
pounds
(23 kg)
77
pounds
(35 kg)
45¼x7x7 inches
(115x30x30 cm)
- Column
counterweight
178.6
pounds
(81 kg)
189.6
pounds
(86 kg)

Table of Contents