SERVICE MANUAL
Checkout and adjustment
RTG PLUS (120V) (Rev. 0)
7-30
7.7.2
7.7.27.7.2
7.7.2
PAN
PANPAN
PAN-
--
-CHILD
CHILDCHILD
CHILD
In the Pan (27), Child (34) function, the possible adjustments are similar
to those described for the Adult size, but the central area deceleration
function is optional and the activation of such function is carried out by
selecting S9 OFF (see description of S9 function on paragraph 7.2.2).
Also, the value of the string "A2 ......." (X-ray emission start angle) can be
changed from 14° (014.0) to 17° (017.0).