SERVICE MANUAL
Checkout and adjustments
(Rev. 0) RTG PLUS (120V)
7-45
7.10
7.107.10
7.10
System reconfiguration after testings
System reconfiguration after testingsSystem reconfiguration after testings
System reconfiguration after testings
After checking the various modes, set them (SO-SA) in the starting
conditions, i.e. in the conditions reported in the “Final Test Table” (see
the last page of the User’s Manual), with the exception of those modes
whose modifications are considered to be necessary.
In case of changes respect to the values reported on the User’s manual,
the new setted values have to be written on the above mentioned table,
in order to assure traceability of these values in case of data corruption
and/or hand control replacement. These values has to be written by the
technical personnel, together with the date of change.