E • Technical Information | Dexcom G7 User Guide 157
Model G7 sensor/transmitter Dexcom receiver
Modulation Gaussian Frequency-Shift Keying
Data Rate 1Mbps
Data
Communication
Range
20feet
The maximum surface temperature of Applied part = 109.4°F.
Essential performance
The G7 CGM system measures patients' glucose sensor readings with specified
accuracy under the stated operating conditions. The Essential Performance of the G7
CGM system also includes reporting the corresponding measured glucose sensor
readings and alerts on the display device.
Quality of service summary
Quality of Service for the G7 System wireless communication using Bluetooth Low
Energy is assured within the effective range of 20feet, unobstructed, between the G7
transmitter and paired display device at regular 5-minute intervals. If connection is
lost between the transmitter and display device, upon re-connection any missed
packets (up to 24hours) will be transmitted from the transmitter to the display device.
The G7 CGM System is designed to only accept radio frequency (RF) communications
from recognized and paired display devices.