EasyManua.ls Logo

Dionex ASE 150 - B.2.8 Adjusting the Cell Holder

Dionex ASE 150
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ASE 150 Operator’s Manual
116 Doc. 065207-02 9/08
B.2.8 Adjusting the Cell Holder
When the ASE 150 is shipped from the factory, the cell holder is
positioned to hold a 100-mL sample cell (or a long rinse cell). To
accommodate a cell of any other size, move the lower latch of the cell
holder (see Figure B-7
) to the appropriate mounting site on the cell door
(see Figure B-8
).
1. Push the
OPEN lever down and open the cell door.
2. Use a 3-mm hex screwdriver (P/N 060154) to remove the two hex
screws (P/N 046253) securing the lower latch to the cell door (see
Figure B-7
).
If the ASE 150 has recently finished a run, the inside of the cell door
will be hot. Put on thermal gloves (P/N 060372) before adjusting the
cell holder.
Si l’ASE 150 a récemment fini l’extraction, la température du four sera
élevée. Utilisez des gants de protection thermique (P/N 060372) avant
de toucher le four ou la cellule d’extraction.
Wenn der ASE 150 gerade einen Lauf beendet hat, ist die Zellentür auf
der Innenseite heiß. Tragen Sie daher Handschuhe (P/N 060372), wenn
Sie den Zellenhalter justieren.
Figure B-7. Removing the Cell Holder Lower Latch
Tab on
underside of
lower latch
Remove two
screws to
change the cell
holder position

Table of Contents