ASE 150 Operator’s Manual
128 Doc. 065207-02 9/08
C.1.3 Setup Screen
Use the SETUP screen to specify the method to run, the sample cell size,
and other ASE 150 operational parameters. All parameters do not fit on
one screen. To view additional parameters, press the down arrow button.
Figure C-4. Setup Screen (All Parameters Shown)
Parameter Description
METHOD
Specifies the name or number of the method to run.
CELL SIZE
Specifies the sample cell installed (1 mL, 5 mL, 10 mL,
22 mL, 34 mL, 66 mL, or 100 mL). The ASE 150
determines various internal operating parameters (for
example, the rinse volume) based on the selected cell
size.
UNITS
Specifies the unit of measure for pressure displays (psi,
MPa, bar, or atm).
REDUCE RELIEF
Determines whether the pressure relief valve is open
for a reduced period (
ON) or the normal period (OFF).
When the pressure relief valve is open, residual
pressure is released from the sample cell. The default
setting (
OFF) is recommended.
PREHEAT TIME
Specifies for how long the oven heats up without
solvent (0 to 99 minutes). The default setting (
0) is
recommended.
PREHEAT PURGE
Enables and disables purging of the cell (without
solvent) during the preheat period. The default setting
(
OFF) is recommended.
METHOD
CELL SIZE
UNITS
OFF
PSI
100ML
PCB
REDUCE RELIEF
PREHEAT TIME
PREHEAT PURGE
BYPASS HEATUP
OFF
OFF
0MIN
KEY SOUND
ON
ERROR SOUND ON
Press the down
arrow button to
scroll through the
list of parameters.