2 Description
Doc. 065207-02 9/08 19
2.4 Method Control
A method determines how the ASE 150 performs the sample extraction. Methods
are defined on the
METHOD EDITOR screen (see Figure 2-6).
The following method parameters define the method. See Section C.1.4
for
detailed descriptions of the method parameters.
TEMPERATURE—the temperature at which to heat the cell.
STATIC TIME—the number of minutes to maintain the cell contents (sample
and solvent) at the temperature set point.
RINSE VOLUME—the amount of solvent to rinse through the sample cell after
the static extraction.
PURGE TIME—the amount of time the cell is purged with nitrogen.
STATIC CYCLE—the number of times the static extraction and rinse cycles are
performed.
CELL TYPE—the type of cell to use (stainless steel or zirconium).
Two types of methods are available:
Preprogrammed methods are application-specific methods created by Dionex.
These methods cannot be changed or deleted by the user. For a list of the
parameters for each preprogrammed method, see Section 2.5
.
Custom methods are user-programmable methods. All custom methods
initially contain the default method parameters. You can modify these
methods as required. For instructions on how to edit a custom method, see
Section 3.6
.
Figure 2-6. Method Editor Screen (Initial View)
TEMPERATURE
METHOD
STATIC TIME
60%
5MIN
100C
1
RINSE VOLUME