C.2.5 Extraction Counters Screen
Use the EXTRACTION COUNTERS screen to check the number of
extractions the system has run or the number of pump strokes performed,
and to reset the counters. This can be helpful in scheduling the routine
replacement of certain parts.
Figure C-10. Extraction Counters Screen
Parameter Description
EXTRACTION
Indicates the number of extractions performed since
the ASE 150 was installed or the extraction counter
was last reset.
PUMP STROKE
Indicates the number of pump strokes performed
since the ASE 150 was installed or the pump stroke
counter was last reset.
EXTR. RESET
Select RESET to reset the extraction counter to zero.
Reset this counter after rebuilding the static valve
(see Section 5.8
).
PUMP RESET
Select RESET to reset the pump stroke counter to
zero. Reset this counter after replacing the pump
seals (see Section 5.6
).
EXTRACTION
PUMP STROKE
EXTR. RESET
PUMP RESET