2
Working with these Instructions
for Use
Header line – the title...
of the main chapter
The title of the specific sub-section is printed underneath the
main header – to help you find your way quickly from subject to
subject.
Page body
...
the Ins
tructions f
or Use
in combined text/illus
trations. The inf
ormation is expressed in
the form of practical actions, giving the user direct hands-on
experience in lear
ning how to use the machine.
Left-hand column...
the t
e
xt
provides explanations and ins
tructs the user s
tep-by-step in the
practical use of the product, with short, clear ins
tructions in
easy-to-follow sequence.
Bullet points indicate separate actions. Where several actions
are described, numbers are used both to refer to the relevant
details in the illustrations and to specify the sequence of
actions.
Right-hand column...
the illus
trations
provide the visual ref
erence f
or the t
ext and make it easier to
locate the various parts of the equipment. Elements mentioned
in the t
ext are highlighted. Unnecessary details are avoided.
Screen displays prompt the user to proceed and confirm
correct actions.
Preparing for use
18
Fitting components
Preparing for use
The following instructions cover:
Assembly of the various equipment items.
Electrical connection and gas connection.
Setting the language for display texts.
Automatic device check with calibration of the sensors.
Fitting components
Always use properly prepared parts, see "Care", page 90.
Fitting the expiration valve
Push the expiration valve block firmly into the mounting.
Check that it is properly secured by gently pulling the port.
Fitting the flow sensor
1 Push the socket to the left as far as it will go.
2 Insert the flow sensor – with the probe facing towards the
ventilator – into the mounting and push it into the socket as
far as it will go.
Then:
3 Push flow sensor to the right as far as it will go into the
rubber lip of the expiration valve.
052053
1
054
2
3
000

Table of Contents

Other manuals for Dräger Evita 2 Dura

Related product manuals