Supplement – Infinity
®
Delta Series Software VF10.1 95
Supplement for Infinity Delta Series Software
Invasive Blood Pressure
Mode of operation Continuous
Protection against electric
shock
Sensor: Type BF
Defibrillator protection Yes
Environmental requirements
Temperature range Operating: 10 to 40 °C (50 to 104 °F)
Storage: –40 to 75 °C (–40 to 167 °F)
Relative humidity Operating: 20 to 90%, non-condensing
Storage: 10 to 95% (with packaging)
Atmospheric pressure Operating: 525 to 795 mmHg (70 to 106 kPa)
Storage: 86 to 825 mmHg (11.5 to 110 kPa)
Physical specifications
Frequency ranges DC to 8, 16, or 32 Hz ± 10% (user selectable)
Accuracy ±1 mmHg or ±1% exclusive of transducer (whichever is greater)