71
The menu » Settings «
– Warning interval
This function is used to set the interval between the warning signals » « on the
analogue interface.
Select the menu items » Settings «, » Communication «, » Analogue interface «
and » Warning interval « in this order, pressing the » « key after each selection.
Use the » « and » « keys to set each position of the time interval (in sec-
onds) for T
1
and T
2
and press the » « key to confirm the setting.
Press the » « key to confirm the complete time interval.
– Warning level
This function is used to set the current on the analogue interface for the warning
signal » «.
Select the menu items » Settings «, » Communication «, » Analogue interface «
and » Warning level « in this order, pressing the » « key after each selection.
Use the » « and » « keys to set each position of the current value and press
the » « key to confirm the setting.
Press the » « key to confirm the complete current value.
– Maintenance signal
This function is used to set the signal type on the analogue interface for the
maintenance signal » «.
Select the menu items » Settings «, » Communication «, » Analogue interface «
and » Maintenance signal « in this order, pressing the » « key after each
selection.
Select the » static « or » dynamic signal type, as required, and activate it by
pressing the » « key.
OK
M
OK
05523758_1_en.eps
t
[s]
T
1
4
20
[mA]
T
2
3
Measured
value
OK
OK
M
OK
OK
OK
OK