The menu » Settings «
72
Note:
The » static « signal type is a DC current signal whose current value can be
configured (see » Maintenance level «).
The » dynamic « signal type is an AC signal with the following characteristic:
– Maintenance level
This function is used to set the current on the analogue interface for the mainte-
nance signal » «.
Select the menu items » Settings «, » Communication «, » Analogue interface «
and » Warning level « in this order, pressing the » « key after each selection.
Use the » « and » « keys to set each position of the current value and press
the » « key to confirm the setting.
Press the » « key to confirm the complete current value.
05923758_1_de.eps
t
[s]
1,1
[mA]
0,7
5
3
OK
M
OK
OK