77
The menu » Settings «
Sensor configuration
(using an EC–O
2
sensor as an example):
This group contains the setting functions for the sensor.
– Gas setting
This function is used to change the settings for the gas to be measured.
Select the menu items » Settings «, » Sensor « , « EC–O
2
« and » Set gas « in this
order, pressing the » « key after each selection.
When several measuring gases of the sensor are displayed:
Use the » « and » « keys to select from the list of measuring gases and the
» « key to activate.
The selected gas is indicated.
Use the » « and » « keys to select the unit of measurement (e.g. Vol.%)
from the list and press the » « key to activate.
The overview of the settings are displayed.
If the settings are correct:
Use the » « key to select the line » Confirm « and press the » « key.
– Sensor initialisation
This function is used to set all parameters of the sensors back to the factory de-
fault values.
Select the menu items » Settings «, » Sensor «, » EC–O
2
« and » Sensor init. «
in this order, pressing the » « key after each selection.
Select » Confirm « and press the » « key to initialise the sensor.
– Set calibration interval
This function is used to set the calibration interval.
Select the menu items » Settings «, » Sensor «, » EC–O
2
« and » Set cal. Int. «
in this order, pressing the » « key after each selection.
Use the » « and » « keys to set each position of the calibration interval
(in days) and press the » « key to confirm the setting.
To confirm the entered calibration interval:
Select » Confirm « and press the » « key.
OK
M
OK
M
OK
M
OK
OK
OK
OK
M
OK
OK