Checkpoint is it a quickly to the calibration state verification. It is only used after the
calibration process. This should only be used after it has been calibrated at least
once. The patient should look at the red light (Led) that lights inside the goggles.
Verification checks if the mask has been moved, in which case it is necessary to re-
calibrate again.
In addition to the position and head movements assistance, in this
window a frame is shown whose color (green: Good, yellow: regular, red:
insufficient) indicates the state of detection of pupil according to the
count, lighting.
If the state is red, it is recommended to reduce the ambient lighting and / or to use
the mask cover if possible. Iftheconditionpersists, contacttechnicalservice.
Toolbarbottom
Home: Return to main window
Back: Return to previous window
Go to: SelectanotherWindow
Next: Forward to next window
HEAD CALIBRATION
Toolbarbottom
Home: Return to main window
Back: Return to previous window