SECURITY WARNINGS
Before using VNG plus with a patient, familiarize yourself with this user manual,
especially with regard to the safety considerations indicated. This manual is only a
reference; its purpose is not to replace its protocol for the safe use of
videonystagmography
This manual uses the following conventions:
WARNING: Alerts the user of the possibility of injury or other serious adverse
reactions associated with the use or misuse of the device
CAUTION: Alerta al usuario de la posibilidad de un problema conel dispositivo
relacionado con su uso o mal uso, como unfuncionamiento incorrecto, un fallo o
daños en el dispositivoo daños a otros bienes.
NOTE: Emphasize information of special importance
.
WARNINGS
• To avoid the risk of electric shock, it must be connected to an AC power
supply with protective grounding.
• To reduce the risk of fire, use the equipment in well-ventilated areas and / or
in the presence of potentially explosive or flammable materials as
environments rich in oxygen or the presence of flammable anesthetic gases.
• The equipment may cause radio interference or may affect the operation of
nearby equipment. It may be necessary to reorient or relocate it.
• Connect it to a power supply network with the specified voltage (100-230V,
50-60 Hz)
• All non-medical equipment that connects to the VNG Plus, must comply with
electrical safety standards according to IEC or ISO of application to said
device. Also, all configurations must comply with the requirements for
medical electrical systems (See IEC 60601-1 Ed. 3.1) When connecting
other equipment to medical electrical equipment, a medical system is being
configured and, consequently, is responsible for ensuring that the system
meet the requirements for medical electrical systems. Keep in mind that local
laws may take precedence over the aforementioned requirements.
• Do not install the equipment near heat sources, such as stoves.
• Install the equipment on firm and horizontal surfaces.
• To reduce the risk of the device heating up, do not obstruct the ventilation
grilles so that there is good air circulation.
• Do not use cables, adapters or extenders that have not been approved by
Ecleris.
• To reduce the risk of electric shock, regularly inspect the AC power cord and
check that it has not frayed or split.