6
• To reduce the risk of strangulation, dispose of the cables so that they do not
get tangled
• To avoid possible injury to the patient due to the use of the device, I
evaluated if the patient is apt to perform nystagmographic tests (See:
Consideration of the patient)
• Only doctors of otolaryngologists, neurologists, balance disorders specialists
or properly trained doctors are authorized to operate the system.
• Turn off the switch before connecting / disconnecting accessories.
• Always position the device so that you can see the PC screen properly and
reach it comfortably,
• The mask contains sensitive electronic and optical devices. Do not subject it
to mechanical shocks
• Turn off the equipment and disconnect it from the AC mains before cleaning
or repairing it.
• Equipment modification is not allowed and may affect performance and
performance. Maintenance must be carried out by qualified technical service
personnel.
• To avoid the risk of communication failures, connect the equipment to USB
ports onboard.
CAUTIÓN:
• Do not attempt to sterilize the mask or any other part of the equipment in an
autoclave or other way.
• To avoid possible damage to the equipment, use only the cleaning and
disinfection products and procedures indicated in this manual.
• To avoid damage, do not drip or spray liquids directly on any part of the
equipment.
• The VNGPLUS has been designed to work in certain environmental
conditions. Respect the following conditions:
Operation Transport
Temperature 10°C-40°C -30°C-50°C
Relative Humidity 30-75% 10-95%
Pressure Atmospheric 700-1060 hPa 500-1060 hPa
• Always transfer it in its original packaging
NOTE:
• The images shown in this manual may not exactly match that of your
VNGPLUS