6
To reduce the risk of strangulation, dispose of the cables so that they do not
get tangled
To avoid possible injury to the patient due to the use of the device, I
evaluated if the patient is apt to perform nystagmographic tests (See:
Consideration of the patient)
Only doctors of otolaryngologists, neurologists, balance disorders specialists
or properly trained doctors are authorized to operate the system.
Turn off the switch before connecting / disconnecting accessories.
Always position the device so that you can see the PC screen properly and
reach it comfortably,
The mask contains sensitive electronic and optical devices. Do not subject it
to mechanical shocks
Turn off the equipment and disconnect it from the AC mains before cleaning
or repairing it.
Equipment modification is not allowed and may affect performance and
performance. Maintenance must be carried out by qualified technical service
personnel.
To avoid the risk of communication failures, connect the equipment to USB
ports onboard.
CAUTIÓN:
Do not attempt to sterilize the mask or any other part of the equipment in an
autoclave or other way.
To avoid possible damage to the equipment, use only the cleaning and
disinfection products and procedures indicated in this manual.
To avoid damage, do not drip or spray liquids directly on any part of the
equipment.
The VNGPLUS has been designed to work in certain environmental
conditions. Respect the following conditions:
Operation Transport
Temperature 10°C-40°C -30°C-50°C
Relative Humidity 30-75% 10-95%
Pressure Atmospheric 700-1060 hPa 500-1060 hPa
Always transfer it in its original packaging
NOTE:
The images shown in this manual may not exactly match that of your
VNGPLUS