C3A, C6A, C6A HDVideoColposcope User Manual EMC Information
- 102 -
A4.2 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
TheVideoColposcopeis intended for use in the electromagnetic environment specified
below. The customer or the user oftheVideoColposcope should assure that it is used in
such an environment.
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
2kV for power
supply lines
Not Applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
1 kV line(s) to
line(s)
2 kV line(s) to
earth
1 kV line(s) to
line(s)
2 kV line(s) to
earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Power frequency
(50Hz/60Hz)
magnetic field
IEC61000-4-8
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.