EasyManua.ls Logo

EDAN C3A - A4.3 Electromagnetic Immunity

EDAN C3A
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C3A, C6A, C6A HDVideoColposcope User Manual EMC Information
- 104 -
A4.3 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
The Video Colposcopeis intended for use in the electromagnetic environment specified
below. The customer or the user oftheVideoColposcope should assure that it is used in
such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Video Colposcope including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
Conducted
RF
IEC
61000-4-6
3 V
rms
150 kHz to 80
MHz
6Vrms
c)
in ISM
bands between
0,15 MHz and 80
MHz
3 V
rms
150 kHz to
80 MHz
6Vrms
c)
in
ISM bands
between
0,15 MHz
and 80 MHz
Pd 2.1
150 kHz to 80 MHz
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz to 2.7
GHz
3 V/m
80 MHz to
2.7 GHz
Pd 2.1
80 MHz to 800 MHz
Pd 3.2
800 MHz to 2.7 GHz
E/6 Pd
at RF wireless
communications equipment bands
(Portable RF communications
equipment (including peripherals
such as antenna cables and external
antennas) should be used no closer
than 30 cm (12 inches) to any part of
the Video Colposcope, including
cables specified by the

Table of Contents