C3A, C6A, C6A HDVideoColposcope User Manual Maintenance and Cleaning
- 90 -
6.3 Cleaning
WARNING
Make sure that the power is off and the power cords are disconnected prior to cleaning or
disinfectingthedevice.
The device should be kept clean from dust, medical solution and reagent. Clean the device when
necessary.
The recommended cleaning agents are:
Mild near neutral detergent
Ethanol (75%)
Isopropanol (70%)
Cleaning the external surface:
1. Switch off the system and remove the power cord.
2. Wipe the external surface of the device components, camera and cables using a soft cloth
dampened with the cleaning agent.
3. Wipe the surface with a soft dry cloth.
Cleaning the lens:
1. Wipe the dust off the lens with a soft brush.
2. Apply a drop of lens cleanser on a dedicated cleaning cloth, and wipe the lens gently.
3. Clean the lens with a soft dry cloth.
CAUTION
1 The device components are not waterproof. Do not immerse any components in water
or allow liquids to enter the cases.
2 In case of liquid splashing on the equipment, wipe it with dry cloth immediately to stop
the liquid from entering the interior. Stop using the device and contact the service
personnel if accidental wetting occurs.
3 Do not allow any remaining solution on the surface of the device, wipe it clean with a
cloth.
4 Do not use strong solvent, for example, acetone.
5 Never use an abrasive such as steel wool or metal polish.
6 Make sure that a dedicated cleaning cloth is used when cleaning the lens. Common
paper or tissue may scratch the lens.
7 Only dedicated lens cleanser should be used to clean the lens, and it should not be
applied on the lens directly.
6.4 Disinfection
This device does not have direct contact with the patient, and in order to protect the product,