M3A Vital Signs Monitor User Manual Intended Use and Safety Guidance
- 4 -
WARNING
22 The simultaneous use of cardiac pacemaker and other patient-connected equipment
may cause safety hazard.
23 Only NIBP and SpO
2
applied parts of the monitor are defibrillation-proof. When a
defibrillator is applied, keep other accessories away from the patient. Otherwise, it
may result in damaging the monitor or harming the patient.
24 Do not touch the patient, table or instrument during defibrillation.
25 During monitoring, if the power supply is off and there is no battery for standby, the
monitor will be off. All last settings used will be recovered when the power is restored.
26 The equipment can provide protective means to prevent the patient from being
burned when used with HF SURGICAL EQUIPMENT. Use only EDAN-approved
accessories.
27 The appliance coupler or mains plug is used as isolation means from supply mains.
Position the monitor in a location where the operator can easily access the
disconnection device.
28 Do not service or maintain the monitor or any accessory which is in use with the
patient.
29 Additional multiple socket-outlet or extension cord cant be connected to the system.
30 Measured data (including trend data, review data and so on) are saved in dataflash
either when the monitor is turned off or when the monitor is powered down in the
process of monitoring.
31 Only items that have been specified as part of the system or specified as being
compatible with the system can be connected to the system.
32 SHOCK HAZARD - Don't connect electrical equipment, which has been supplied as a
part of the system, directly to the wall outlet when the non-medical equipment is
intended to be supplied by a multiple portable socket-outlet with an isolation
transformer.
33 SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
34 All the accessories connected to system must be installed outside the patient vicinity,
if they do not meet the requirement of IEC/EN 60601-1.
35 Using accessories other than those specified may result in increased electromagnetic
emission or decreased electromagnetic immunity of the monitoring equipment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN M3A and is the answer not in the manual?

EDAN M3A Specifications

General IconGeneral
TypeVital Signs Monitor
SpO2 Measurement Range0-100%
ParametersSpO2, Pulse Rate, Temperature
Pulse Rate Measurement Range30 to 250 bpm
Temperature Measurement Range0 to 50°C
Power SupplyAC 100-240V, 50/60Hz
Battery LifeUp to 8 hours