M3A Vital Signs Monitor User Manual System Menu
- 28 -
NOTE:
1 If data storage reaches the maximum, the monitor cant store data; the user needs to
manually delete some data to continue.
2 It is recommended to format the USB flash drive to the FAT file type via PC prior to
use.
Please refer to Chapter 6.1 Trend List for details.
4.7 Recorder
Select Recorder in Main Menu to open the menu.
Record Realtime Data: Select it to output the real time data from the monitor.
Record Trend Graph: Select it to output the trend graph.
Record Current Trend Table: Select it to output the current trend table.
Record History Trend Table: Select it to output the history trend table.
Please refer to Chapter 6 Recording for more information.
4.8 System Setup
There are a few items to be set in this submenu: General Setup, General Alarm Setup,
Time&Date Setup, and Default Configuration.
4.8.1 General Setup
Select General Setup in System Setup to open submenu as below:
Key Volume: set key volume to level 0 ~ 5.
LCD Brightness: set LCD brightness to level 1 ~ 5.
Standby Mode: set to ON or OFF. If you set this item to ON, when pressing ON/OFF
button for less than 1s, the monitor will enter Standby Mode. (Please refer to 4.10 Standby
Mode for more information.)
Work Mode: set to Spot or Monitor.
4.8.2 General Alarm Setup
Select General Setup in System Setup to open submenu.
Alarm Volume: set alarm volume to level 1 ~ 3.
4.8.3 Time & Date Setup
Select Time & Date Setup in System Setup to access the submenu. System time is in format of
yy-mm-dd, mm-dd-yy or dd-mm-yy. Users can set the year, month, day, hour, minute and
second. Select the item you want to modify and confirm it by pressing Confirm. Select Cancel

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN M3A and is the answer not in the manual?

EDAN M3A Specifications

General IconGeneral
TypeVital Signs Monitor
SpO2 Measurement Range0-100%
ParametersSpO2, Pulse Rate, Temperature
Pulse Rate Measurement Range30 to 250 bpm
Temperature Measurement Range0 to 50°C
Power SupplyAC 100-240V, 50/60Hz
Battery LifeUp to 8 hours