M3A Vital Signs Monitor User Manual Alarm
- 38 -
Other audible signals are not affected, including beat volume, key volume.
Pressing the button again can resume the audio alarm.
When Audio Alarm Off is set to OFF, this function is inactivated. The monitor has no response
if you press the button .
NOTE:
If a new alarm occurs during the audio alarm off period, the new alarm will not be
sounding.
5.5 Alarm Reset
To reset the alarm, you can select Main Menu > Alarm Setup > Alarm Reset. After resetting
the alarm,
The audio alarm (including technical alarm and physiological alarm) is turned off, and no
alarms are sounding.
The visual alarm indications are still displayed.
It will not influence the configuration of physiological alarm off, audio alarm paused and
audio alarm off status.
NOTE:
If a new alarm occurs after the alarm is reset, the new alarm will be sounding.
5.6 Parameter Alarm
WARNING
1 Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
2 Setting alarm limits to extreme values may cause the alarm system to become
ineffective.
In Main Menu > Alarm Setup, you can check and set the alarm limit or alarm status. The setup
is isolated from each other.
When a parameter alarm is OFF, an icon displays near the parameter. If the alarms are
turned off individually, they must be turned on individually.
For the parameters whose alarm is set to ON, the alarm will be triggered when at least one of
them exceeds alarm limits. The following actions take place:
1. Alarm message displays on the screen as described in alarm mode;
2. The monitor beeps in its corresponding alarm class and volume;
3. Alarm lamp flashes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN M3A and is the answer not in the manual?

EDAN M3A Specifications

General IconGeneral
TypeVital Signs Monitor
SpO2 Measurement Range0-100%
ParametersSpO2, Pulse Rate, Temperature
Pulse Rate Measurement Range30 to 250 bpm
Temperature Measurement Range0 to 50°C
Power SupplyAC 100-240V, 50/60Hz
Battery LifeUp to 8 hours

Related product manuals