M3A Vital Signs Monitor User Manual NIBP Monitoring (Optional)
- 60 -
4. Wait until the first reading is taken.
NOTE:
1 The width of the cuff is either approximately 40% of the limb circumference or 2/3 of
the upper arm length. The inflatable part of the cuff should be long enough to encircle
80%-100% of the limb. The wrong size of cuff can cause erroneous readings. If the
cuff size is in question, use another cuff with suitable size to avoid errors.
2 If an NIBP measurement is suspect, repeat the measurement. If you are still
uncertain about the reading, use another method to measure the blood pressure.
3 Please make sure the cuff is well connected. A leak of air may cause measurement
error.
4 Please select the cuff with the suitable size. An unsuitable cuff may cause incorrect
measurements.
5 Avoid incursion of liquid into the cuff. If this happens, please desiccate the cuff
completely.
9.4.1 Operation Prompts
To start auto measurement
Access NIBP Setup menu and pick the Interval item, in which the user may set up the time
interval for auto measurement. After that, press the NIBP START/STOP on the front panel
to start the auto measurement according to the selected time interval.
To stop auto measuring:
During auto measuring, press the NIBP START/STOP on the front panel at any time to stop
auto measurement in process.
WARNING
If you repeatedly use AUTO measuring in a short term, it may lead to inaccurate readings
or endanger patient’s life.
To start a manual measuring:
Access NIBP Setup menu and pick the Interval item. Select the Manual selection.
Then press the NIBP START/STOP on the front panel to start a manual measurement.
WARNING
Prolonged NIBP measurements in Auto mode may be associated with purport, ischemia
and neuropathy in the limb wearing the cuff. When monitoring a patient, examine the
extremities of the limb frequently for normal color, warmth and sensitivity. If any
abnormality is observed, stop the blood pressure measurements.
To start a continuous measuring:
Access the NIBP Setup menu and pick the Continual item to start a continuous
measurement. The continuous measurement will last 5 min.